DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 2: Performance verification
DMM6500-905-01 Rev. E June 2021
2-11
To verify ac voltage accuracy:
1. Connect the DMM6500 HI and LO INPUT connectors to the calibrator as shown in the following
figure.
Figure 2: Connections for ac voltage verification 100 mV to 100 V ranges
2. On the DMM6500, press the
FUNCTION
key and select
AC voltage
.
3. On the home screen, select the button next to Range and select
100 mV
.
4. Press the
MENU
key.
5. Under Measure, select
Settings
.
6. Make sure that detector bandwidth is set to
30 Hz
.
AC voltage is specified for the detector bandwidth setting of 3 Hz. Three Hz measures accurately for
input signals from 3 Hz to 300
kHz, with reading rates ≈ 2 readings/s. T
o improve verification
throughput to ≈ 20 readings/s, set detector bandwidth to 30
Hz for frequencies of 30 Hz to 300 kHz.
To verify frequencies 1 kHz and higher, set the detector bandwidth to 300
Hz for faster ≈ 200
readings/s throughput.
7. Source ac voltages for each of the frequencies listed in the
Verify the ac voltage 100 mV range
(on page 2-12) table.
8. Repeat these steps for each range and frequency listed in the tables below. For each voltage
setting, be sure that the reading is within low and high limits.
Summary of Contents for DMM6500
Page 1: ...6 Digit Multimeter with Scanning Calibration and Adjustment Manual DMM6500...
Page 5: ......
Page 6: ......
Page 64: ......