Calibration
2-19
Resistance calibration
Follow these steps for resistance calibration:
1.
Set the calibrator to the resistance mode, and turn on external sensing.
NOTE
Use external sense (4-wire
Ω
) when calibrating all resistance ranges. Be sure that the
calibrator external sense mode is turned on.
2.
Perform the calibration steps summarized in Table 2-10. For each step:
•
Set the calibrator to the indicated resistance, and make sure the unit is in operate.
(Use the recommended resistance or the closest available value.)
•
Send the indicated programming command. (Change the command parameter if
you are using a different calibration resistance than that shown.)
•
Wait until the Model 2015 completes each step before continuing.
Table 2-10
Resistance calibration programming steps
Calibration
step
Calibrator
resistance
Calibration command*
Parameter range
1k
Ω
10k
Ω
100k
Ω
1M
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
:CAL:PROT:DC:STEP6 1E3
:CAL:PROT:DC:STEP7 10E3
:CAL:PROT:DC:STEP8 100E3
:CAL:PROT:DC:STEP9 1E6
900 to 1.1E3
9E3 to 11E3
90E3 to 110E3
900E3 to 1.1E6
* Use exact calibrator resistance value for parameter.
Summary of Contents for 2015
Page 15: ...1 Performance Verification...
Page 34: ...2 Calibration...
Page 60: ...3 Routine Maintenance...
Page 64: ...4 Troubleshooting...
Page 89: ...5 Disassembly...
Page 100: ...5 12 Disassembly...
Page 101: ...6 Replaceable Parts...
Page 115: ...A Specifications...
Page 127: ...B Calibration Reference...
Page 149: ...C Calibration Program...
Page 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...