Troubleshooting
4-21
Analog signal switching states
Tables 4-5 through 4-11 provide switching states of the various relays, FETs, and analog
switches for the basic measurement functions and ranges. These tables can be used to assist in
tracing an analog signal from the input to the A/D multiplexer.
Table 4-5
DCV signal switching
Table 4-6
ACV and FREQ signal switching
Range
Q101
Q102
Q114
Q136
Q109
K101* Q113
Q105
Q104
Q108
Q121
100mV
1V
10V
100V
1000V
ON
ON
ON
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
OFF
OFF
SET
SET
SET
SET
SET
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
ON
ON
ON
* K101 set states:
Pin 8 switched to Pin 7
Pin 3 switched to Pin 4
Range Q101
Q102
K101* K102*
U103
pin 8
U103
pin 9
U105
pin 9
U105
pin 8
U103
pin 16
U103
pin 1
U105
pin 1
U111
pin 16
100mV
1V
10V
100V
750V
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
RESET
RESET
RESET
RESET
RESET
RESET
RESET
SET
SET
SET
ON
ON
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
OFF
ON
OFF
ON
OFF
OFF
OFF
OFF
OFF
OFF
OFF
* K101 and K102 reset states:
K101 and K102 set states:
Pin 8 switched to Pin 9
Pin 3 switched to Pin 2
Pin 8 switched to Pin 7
Pin 3 switched to Pin 4
Summary of Contents for 2015
Page 15: ...1 Performance Verification...
Page 34: ...2 Calibration...
Page 60: ...3 Routine Maintenance...
Page 64: ...4 Troubleshooting...
Page 89: ...5 Disassembly...
Page 100: ...5 12 Disassembly...
Page 101: ...6 Replaceable Parts...
Page 115: ...A Specifications...
Page 127: ...B Calibration Reference...
Page 149: ...C Calibration Program...
Page 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...