B-14
Calibration Reference
AC calibration commands
The :AC commands perform comprehensive (user) calibration of the ACV and ACI functions.
Table B-3 summarizes these calibration commands.
Table B-3
AC calibration commands
Command
Description
:CALibration
:PROTected
:AC
:STEP1
10mV AC at 1kHz calibration step.
:STEP2
100mV AC at 1kHz calibration step.
:STEP3
100mV AC at 50kHz calibration step.
:STEP4
1V AC at 1kHz calibration step.
:STEP5
1V AC at 50kHz calibration step.
:STEP6
10V AC at 1kHz calibration step.
:STEP7
10V AC at 50kHz calibration step.
:STEP8
100V AC at 1kHz calibration step.
:STEP9
100V AC at 50kHz calibration step.
:STEP10
700V AC at 1kHz calibration step.
:STEP11
100mA AC at 1kHz calibration step.
:STEP12
1A AC at 1kHz calibration step.
:STEP13
2A AC at 1kHz calibration step.