Troubleshooting
4-23
Table 4-10
DCA signal switching
Table 4-11
ACA signal switching
Tables 4-12 through 4-16 can be used to trace the analog signal through the A/D multiplexer
(U163) to the final amplifier stage. These tables show the MUX lines (S3, S4, S6, S7) that are
selected for measurement during the SIGNAL phase of the multiplexing cycle. Also included
are switching states of analog switches (U129) that set up the gain for the final amplifier stage
(U166).
Table 4-12
DCV signal multiplexing and gain
Table 4-13
ACV and ACA signal multiplexing and gain
Range
K103
10mA
100mA
1A
3A
ON
OFF
OFF
OFF
Range
K103
U105
pin 16
U105
pin 1
U111
pin 16
U105
pin 8
U103
pin 16
U103
pin 1
1A
3A
OFF
OFF
ON
ON
ON
ON
OFF
ON
OFF
OFF
OFF
OFF
OFF
OFF
Range
Signal
(U163)
U129
pin 1
U129
pin 8
U129
pin 9
Gain
(U166)
100mV
1V
10V
100V
1000V
S4
S4
S4
S4
S4
OFF
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
OFF
OFF
OFF
×
100
×
10
×
1
×
10
×
1
Range
Signal
(U163)
U129
pin 1
U129
pin 8
U129
pin 9
Gain
(U166)
All
S3
ON
OFF
OFF
×
1
Summary of Contents for 2015
Page 15: ...1 Performance Verification...
Page 34: ...2 Calibration...
Page 60: ...3 Routine Maintenance...
Page 64: ...4 Troubleshooting...
Page 89: ...5 Disassembly...
Page 100: ...5 12 Disassembly...
Page 101: ...6 Replaceable Parts...
Page 115: ...A Specifications...
Page 127: ...B Calibration Reference...
Page 149: ...C Calibration Program...
Page 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...