Calibration Reference
B-17
Manufacturing calibration commands
Three calibration steps are only performed at the factory or when the unit has been repaired:
:CALibration:PROTected:AC:STEP14
1V AC at 3Hz
:CALibration:PROTected:AC:STEP15
1V AC at 1kHz
:CALibration:PROTected:DC:STEP0
Rear terminal short circuit
:AC:STEP<14|15>
(CALibration:PROTected:AC:STEP<14|15>)
Purpose
To program individual AC manufacturing calibration steps.
Format
:cal:prot:ac:step14 <Cal_voltage>
:cal:prot:ac:step15 <Cal_frequency>
Parameters
<Cal_voltage> = 1 [1V nominal]
<Cal_frequency = 1E3 [1kHz nominal]
Description
The
:AC:STEP14
and
:AC:STEP:15
commands program the two manu-
facturing AC calibration steps. The appropriate signal must be connected to
the instrument when programming each step, as summarized by the parame-
ters listed above.
Examples
:CAL:PROT:AC:STEP14 1
Program AC step 14.
:CAL:PROT:AC:STEP15 1E3
Program AC step 15.
:DC:STEP0
(:CALibration:PROTected:DC:STEP0)
Purpose
To perform rear terminal short-circuit calibration.
Format
:cal:prot:dc:step0
Parameter
None
Description
:STEP0
performs the rear short-circuit calibration step in the manufacturing
calibration procedure. Connect a low-thermal short (Model 8610) to the rear
panel input jacks, and select the rear inputs before sending this command.
Example
:CAL:PROT:DC:STEP0
Perform rear short-circuit calibration.
Summary of Contents for 2015
Page 15: ...1 Performance Verification...
Page 34: ...2 Calibration...
Page 60: ...3 Routine Maintenance...
Page 64: ...4 Troubleshooting...
Page 89: ...5 Disassembly...
Page 100: ...5 12 Disassembly...
Page 101: ...6 Replaceable Parts...
Page 115: ...A Specifications...
Page 127: ...B Calibration Reference...
Page 149: ...C Calibration Program...
Page 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...