Specifications A-3
Frequency Sweep Reading Rate
NUMBER OF FREQUENCIES
TIME (seconds)
4
5
0.2
30
1.1
100
3.5
200
6.9
Notes
1. Input signal at full scale.
2. V
IN
≥
20% of range and harmonics >–65dB.
3. Speeds are for default operating conditions (*RST), and display off, auto range off, binary data transfer, trig delay = 0.
4. Typical times: frequencies in 400–4kHz range, binary data transfer, TRIG DELAY = 0, Display OFF, Auto Range OFF. Data returned is THD
measurement plus AC voltage.
GENERATOR CHARACTERISTICS
FREQUENCY RANGE:
10–20kHz.
FREQUENCY RESOLUTION:
0.007Hz.
FREQUENCY ACCURACY:
±(0.015% of r 0.007Hz)
1
.
FREQUENCY TEMPERATURE COEFFICIENT:
<100ppm over operating temperature range.
SOURCE OUTPUT:
Waveform:
Sinewave.
Amplitude Range:
2V rms (50
Ω
and 600
Ω
) or 4V rms (HI Z).
Amplitude Resolution:
0.5mV rms (50
Ω
and 600
Ω
) or 1mV rms (HI Z).
Amplitude Accuracy:
±(0.3% of s 2mV)
1, 4
.
Amplitude Temperature Coefficient:
Typically 0.015%/°C.
Amplitude Flatness:
±0.1dB
1, 4, 5
.
Output Impedance:
50
Ω
± 1
Ω
or 600
Ω
± 10
Ω
, user selectable.
THD:
–64dB
6
.
Noise:
100µV rms
2
.
DC Offset Voltage:
±1.2mV
1
.
INV/PULSE OUTPUT (SINEWAVE MODE):
Frequency:
Same as source output.
Amplitude Range:
2V rms (50
Ω
and 600
Ω
) or 4V rms (HI Z).
Amplitude Resolution:
0.5mV (50
Ω
and 600
Ω
) or 1mV rms (HI Z).
Amplitude Accuracy:
±(2.0% of s 2mV)
1, 4
.
Amplitude Flatness:
±0.1dB
1, 4, 5
.
Output Impedance:
Same as Source Output setting.
THD:
–64dB
6
.
Noise:
100µV rms
2
.
DC Offset Voltage:
±1.1mV typ., ±13mV max.
1
Summary of Contents for 2015
Page 15: ...1 Performance Verification...
Page 34: ...2 Calibration...
Page 60: ...3 Routine Maintenance...
Page 64: ...4 Troubleshooting...
Page 89: ...5 Disassembly...
Page 100: ...5 12 Disassembly...
Page 101: ...6 Replaceable Parts...
Page 115: ...A Specifications...
Page 127: ...B Calibration Reference...
Page 149: ...C Calibration Program...
Page 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...