Specifications A-5
DC OPERATING CHARACTERISTICS
2
FUNCTION
DIGITS
READINGS/s
PLCs
8
DCV (all ranges),
6
1
⁄
2 3, 4
5
10
DCI (all ranges), and
6
1
⁄
2 3, 7
30
1
2W Ohms (<10M range)
6
1
⁄
2 3, 5
50
1
5
1
⁄
2 3, 5
270
0.1
5
1
⁄
2 5
500
0.1
5
1
⁄
2 5
1000
0.04
4
1
⁄
2 5
2000
0.01
DC SYSTEM SPEEDS
2, 6
RANGE CHANGE
3
:
50 / s.
FUNCTION CHANGE
3
:
45 / s.
AUTORANGE TIME
3, 10
:
<30 ms.
ASCII READINGS TO RS-232 (19.2K BAUD):
55 / s.
MAX. INTERNAL TRIGGER RATE:
2000 / s.
MAX. EXTERNAL TRIGGER RATE:
400 / s.
DC GENERAL
LINEARITY OF 10VDC RANGE:
±(2ppm of r 1ppm of range).
DCV,
Ω
, TEMPERATURE, CONTINUITY, DIODE TEST INPUT PROTECTION:
1000V, all ranges.
MAXIMUM 4W
Ω
LEAD RESISTANCE:
10% of range per lead for 100
Ω
and 1k
Ω
ranges; 1k
Ω
per lead for all other ranges.
DC CURRENT INPUT PROTECTION:
3A, 250V fuse.
SHUNT RESISTOR:
0.1
Ω
for 3A, 1A and 100mA ranges. 10
Ω
for 10mA range.
CONTINUITY THRESHOLD:
Adjustable 1
Ω
to 1000
Ω
.
AUTOZERO OFF ERROR:
Add ±(2ppm of range error + 5µV) for <10 minutes and ±1°C change.
OVERRANGE:
120% of range except on 1000V, 3A and Diode.
SPEED AND NOISE REJECTION
RMS NOISE
RATE
READINGS/S
DIGITS
10V RANGE
NMRR
12
CMRR
13
10 PLC
5
6
1
⁄
2
< 1.5 µV
60 dB
140 dB
1 PLC
50
6
1
⁄
2
< 4 µV
60 dB
140 dB
0.1 PLC
500
5
1
⁄
2
< 22 µV
—
80 dB
0.01 PLC
2000
4
1
⁄
2
< 150 µV
—
80 dB
DC Notes
1. Add the following to ppm of range accuracy specification based on range:1V and 100V, 2ppm; 100mV, 15ppm; 100
Ω
, 15ppm; <1M
Ω
, 2ppm; 10mA and 1A, 2ppm;
100mA, 20ppm.
2. Speeds are for 60 Hz operation using factory default operating conditions (*RST). Autorange off, Display off, Trigger delay = 0.
3. Speeds include measurement and binary data transfer out the GPIB.
4. Auto zero off.
5. Sample count = 1024, auto zero off.
6. Auto zero off, NPLC = 0.01.
7. Ohms = 24 readings/second.
8. 1 PLC = 16.67ms @ 60Hz, 20ms @ 50Hz/400Hz. The frequency is automatically determined at power up.
9. For signal levels >500V, add 0.02ppm/V uncertainty for the portion exceeding 500V.
10. Add 120ms for ohms.
11. Must have 10% matching of lead resistance in Input HI and LO.
12. For line frequency ±0.1%.
13. For 1k
Ω
unbalance in LO lead.
14. Relative to calibration accuracy.
15. Specifications are for 4-wire ohms or 2-wire ohms with REL function.
Summary of Contents for 2015
Page 15: ...1 Performance Verification...
Page 34: ...2 Calibration...
Page 60: ...3 Routine Maintenance...
Page 64: ...4 Troubleshooting...
Page 89: ...5 Disassembly...
Page 100: ...5 12 Disassembly...
Page 101: ...6 Replaceable Parts...
Page 115: ...A Specifications...
Page 127: ...B Calibration Reference...
Page 149: ...C Calibration Program...
Page 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...