A-2
Specifications
DISTORTION CHARACTERISTICS
VOLTAGE RANGE:
100mV, 1V, 10V, 100V, 750V (user selectable).
INPUT IMPEDANCE:
1M
Ω
paralleled by <100pF.
DISPLAY RANGE:
0–100% or 0–100.00dB.
RESOLUTION:
0.0001% or 0.00001dB.
FUNDAMENTAL FREQUENCY RANGE:
20Hz–20kHz.
HARMONIC FREQUENCY RANGE:
40Hz–50kHz.
FREQUENCY RESOLUTION:
0.008Hz.
FREQUENCY ACCURACY:
±0.01% of reading.
FREQUENCY TEMPERATURE COEFFICIENT:
≤
100ppm over operating temperature range.
MEASUREMENT
ACCURAC Y
RESIDUAL
MODE
(1 Year, 23°C ±5°C)
DISTORTION
1
THD and individual
±0.8dB,
0.004% or –87dB
harmonic magnitudes
20Hz to 20kHz
2
20Hz to 20kHz
THD + n
±1.5 dB,
0.056% or –65dB
100Hz to 20kHz
2
20Hz to 20kHz
SINAD
±1.5dB
+65dB
100Hz to 20kHz
2
20Hz to 20kHz
AC Level
±(0.13% of r
V rms
0.009% of range)
20Hz to 20kHz
Distortion Measurement Audio Filters
None
C-Message
CCITT Weighting
CCIR/ARM
CCIR
“A” Weighting
NUMBER OF HARMONICS INCLUDED IN THD CALCULATION:
2 to 64 (user selectable).
HI AND LO CUTOFF FILTERS (bus settable):
20Hz–50kHz. Can be combined to form brickwall bandpass filter.
Distortion Measurement Reading Rate
3
FUNDAMENTAL
MINIMUM
FREQUENCY
FUNDAMENTAL
READINGS
ACQUISITION
FREQUENCY
PER
MODE
RANGE
SECOND
Single acquisition
20 Hz to 100 Hz
14
or stored value
100 Hz to
1 kHz
24
1 kHz to 20 kHz
28
Automatic
20 Hz to 30 Hz
5.5
30 Hz to 400 Hz
6
400 Hz to 20 kHz
6.6
Summary of Contents for 2015
Page 15: ...1 Performance Verification...
Page 34: ...2 Calibration...
Page 60: ...3 Routine Maintenance...
Page 64: ...4 Troubleshooting...
Page 89: ...5 Disassembly...
Page 100: ...5 12 Disassembly...
Page 101: ...6 Replaceable Parts...
Page 115: ...A Specifications...
Page 127: ...B Calibration Reference...
Page 149: ...C Calibration Program...
Page 158: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...