19
Shop Online
CLASSIC 800 Product Manual
Setup & Operation
1. View or Adjust Relay 1 & 2 Energized Mode
Both relays on the CLASSIC 800 have Normally Open and Normally Closed contacts. The terms ‘Open’ and
‘Closed’ refer to the de-energized state of the relays. Normally Open or Normally Closed operation for
Relay 1/
Relay 2
is determined by wiring the contacts accordingly (see Figures 11 & 12).
Note:
Low flow means less Thermal Signal as represented by the CLASSIC 800’s Thermal Signal Bar Graph.
Some configurations may require one or both relays to be energized on high flow, i.e. higher than the
set point.
To View or Adjust the Relay 1 & 2 Energized Modes:
1.
Press
SET
until the
Relay 1
or
Relay 2
LED is on solid and the
Run Mode
LED is off.
2.
The
Thermal Signal Bar Graph
shows the Relay Mode. For Low Flow the first six
(0-25% LED’s)
are on
solid; for High Flow the last six
(75-100% LED’s)
are on solid.
3.
To adjust the Mode, press and release the
Left
or
Right
Arrow keys.
4.
When satisfied with your configuration, press and release
SET
to continue to the next programmable
variable OR press and release
RUN
to return to
Run Mode.
Figure 15 - Relay Energized Mode
Thermal Signal
100%
0
Relay 1
Relay 2
Fault
Set Point 1
Set Point 2
Run Mode
Bypass
RUN
SET
1
2
CLASSIC
™
Thermal Signal
100%
0
Relay 1
Relay 2
Fault
Set Point 1
Set Point 2
Run Mode
Bypass
RUN
SET
1
2
CLASSIC
™
Relay 1 LED
is ON
Set Relay
Mode LOW
Set Relay
Mode HIGH
LOW
FLOW
HIGH
FLOW
Relay 2 LED
is ON