7 Reprocessing steps in accordance with ISO 17664 | 7.6 Sterilisation
7.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / ISO 17665-1
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.
NOTICE
Contact corrosion due to moisture.
Damage to product.
▶
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
The KaVo medical device has a maximum temperature resistance up to 138
℃
(280.4 °F).
Sterilisation parameters:
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilisation processes:
▪ Steriliser with triple pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Steriliser using the gravity method:
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F) or
alternatively
- at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4 °F)
▶
Remove contra-angle handpieces and turbine handpieces from the steam
steriliser immediately after completion of the sterilisation cycle.
▶
Use according to the manufacturer's Instructions for Use.
7.7 Storage
Reprocessed products must be stored appropriately such that they are protec-
ted from germs (as far as possible) and dust, in a dry, dark, cool room.
Note
Comply with the expiry date of the sterilised items.