7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
18 / 30
7 Reprocessing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶
Take suitable personal protective measures.
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
▶
Reprocess the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to reprocessing.
▶
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
▶
Remove the tool from the medical device.
▶
Remove all residual cement, composite or blood immediately.
▶
Do not place in solutions or similar substance.
7.2 Manual Reprocessing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
Clean manually or in a washer disinfector only.
7.2.1 Manual cleaning - external
Required accessories:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush such as a medium hard toothbrush
▶
Brush off under flowing tap water.
7.2.2 Manual cleaning - internal
Validated internal cleaning (removal of residual protein) can be accomplished
with KaVo CLEANspray.
▶
Cover the medical device with the KaVo Cleanpac bag, and place it on the
corresponding care adapter.