4 Start up and shut down | 4.3 Checking the pressures
12 / 30
4.3 Checking the pressures
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion
Premature wear
▶
Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
▶
Insert the test manometer (
Mat. no. 0.411.8731
) between the coupling
and the medical device and check the following pressures:
ð
Drive pressure, drive pressure recommended, return air pressure, spray
water pressure, and spray air pressure.
See also:
2
3.2 Technical Specifications, Page 10
4.4 Checking the O-rings
NOTICE
Missing or damaged O-rings.
Malfunction and premature failure.
▶
Make sure that all O-rings are on the coupling and are undamaged.
Number of available O-rings: 5