5 Operation | 5.1 Attaching the medical device
5 Operation
Note
At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.
5.1 Attaching the medical device
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked in place can release from the MUL-
TIflex (LUX) / MULTIflex LED coupling during treatment.
▶
Before each treatment, check if the medical device is securely locked onto
the MULTIflex (LUX) / MULTIflex LED coupling by pulling on it.
NOTICE
Inaccurate coupling can destroy the high-pressure lamp or the LED of the
coupling or reduce its service life.
▶
Make sure that the turbine is accurately coupled and firmly seated on the
coupling.
▶
Mount the medical device accurately on the MULTIflex (LUX) / MULTIflex
LED coupling and push it backward until the coupling audibly locks in the
medical device.
▶
Pull on it to make sure that the medical device is securely affixed to the
coupling.
5.2 Removing the medical device
▶
Grasping the coupling, twist the medical device slightly and pull it off.
5.3 Inserting the cutter or grinder
Note
Only use carbide cutters or diamond grinders that correspond to DIN EN ISO
1797-1 type 3, are made of steel or hard metal and meet the following cri-
teria:
- Shaft diameter: 1.59 to 1.60 mm (0.0626 in to 0.0629 in)
- Overall length: max. 25 mm (0.9842 in)
- Shaft clamping length: at least 11 mm (0.4330 in)
- Blade diameter: max. 2 mm (0.0787 in)
WARNING
Use of unauthorised dental burs or diamond grinders.
Injury to the patient or damage to the medical device.
▶
Comply with the instructions for use and use the dental bur or diamond
grinder properly.
▶
Only use dental burs or diamond grinders that do not deviate from the spe-
cified data.