background image

Instructions for use MASTERtorque LUX M9000 L COLOR - 1.007.9900 MASTERtorque LUX M9000 L COLOR -
1.007.7100

7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing

20 / 30

Follow the instructions for use of the disinfectant.

7.2.4 Manual disinfection - internal

 WARNING

Incomplete disinfection.

Infection hazard.

Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.

If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.

NOTICE

Never disinfect the handpiece with chloride-containing products.

Malfunction and material damage.

Only disinfect in a washer disinfector or manually.

The efficacy of manual internal disinfection must be demonstrated by the man-
ufacturer of the disinfection agent. With KaVo products, use only disinfection
agents that have been released by KaVo with respect to the compatibility of
materials (e.g. WL-cid / made by ALPRO).

Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.

Hold the can vertically.

Press the spray key for at least 3 seconds.

Remove the medical device from the spray attachment and let the disinfect-
ant act for 2 minutes.

Follow the instructions for use of the disinfectant.

7.2.5 Manual drying

Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts.

Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.

Hold the can vertically.

Press the spray key for at least 3 seconds.

See also:

2

KaVo DRYspray Instructions for Use

Note

KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only
available in the following countries:
Belgium, Denmark, Germany, Finland, France, United Kingdom, Italy, Latvia,
Lithuania, Luxembourg, Netherlands, Norway, Austria, Poland, Portugal,
Sweden, Switzerland, Spain and the Czech Republic.
In other countries interior cleaning can only be carried out with washer disin-
fectors in accordance with EN ISO 15883-1.

Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.

Summary of Contents for MASTERtorque LUX M9000 L COLOR 1.007.9900

Page 1: ...Instructions for use MASTERtorque LUX M9000 L COLOR 1 007 9900 MASTERtorque LUX M9000 L COLOR 1 007 7100...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...n the MULTIflex LUX MULTIflex LED coupling 16 6 2 Cleaning the spray nozzle 16 6 3 Changing the water filter 16 7 Reprocessing steps in accordance with ISO 17664 18 7 1 Preparations at the site of use...

Page 4: ...8 Tools and consumables 26 9 Terms and conditions of warranty 27 Instructions for use MASTERtorque LUX M9000 L COLOR 1 007 9900 MASTERtorque LUX M9000 L COLOR 1 007 7100 Table of contents 4 30...

Page 5: ...e contact your local dealer or the KaVo Repair Service directly 49 0 7351 56 1900 service reparatur kavokerr com Target group This document is intended for dentists and their assistants The startup se...

Page 6: ...re Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal i...

Page 7: ...disposal the product and accessories must be appropriately repro cessed or sterilised In the case of injury to soft tissue do not continue treatment in the oral cavity with compressed air driven inst...

Page 8: ...soft tissue with the handpiece head or instrument cover Do not use the medical device as a light probe Use an appropriate light probe for illumination of the oral cavity or site of preparation After t...

Page 9: ...pplications Removal of carious material cavity and crown preparations removal of fillings processing of tooth and restora tion surfaces A medical device according to relevant national statutory regula...

Page 10: ...essure 1 0 to 2 5 bar 15 to 36 psi Air consumption 42 to 48 Nl min Idle speed 340 000 to 400 000 min 1 Recommended pressure 2 to 3 N Attachable to all MULTIflex LUX MULTIflex LED couplings 3 3 Transpo...

Page 11: ...7494 2 4 1 Mounting the MULTIflex LUX MULTIflex LED coupling WARNING Detachment of the medical device during treatment A medical device that is not properly locked in place can release from the MUL T...

Page 12: ...ean and uncontaminated in accordance with EN ISO 7494 2 Insert the test manometer Mat no 0 411 8731 between the coupling and the medical device and check the following pressures Drive pressure drive p...

Page 13: ...educe its service life Make sure that the turbine is accurately coupled and firmly seated on the coupling Mount the medical device accurately on the MULTIflex LUX MULTIflex LED coupling and push it ba...

Page 14: ...of the service life of tool and chuck system Do not operate the tool at a higher speed than recommended by the manu facturer NOTICE Use of dental burs or diamond grinders with worn or damaged shafts M...

Page 15: ...MASTERtorque LUX M9000 L COLOR 1 007 7100 5 Operation 5 4 Removing the cutter or grinder After the dental bur or diamond grinder has stopped rotating firmly press the press button with your thumb and...

Page 16: ...loop Push the O ring to the front and remove it Insert new O rings into the grooves 6 2 Cleaning the spray nozzle WARNING Hazard from nonsterile products Infection danger to the care provider and pat...

Page 17: ...the tooth Check the filter and exchange if necessary Check the spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 0921 if necessary Unscrew the sleeve in countercloc...

Page 18: ...he risk of infection during reprocessing always wear protect ive gloves Remove the tool from the medical device Remove all residual cement composite or blood immediately Do not place in solutions or s...

Page 19: ...ed out with washer disin fectors in accordance with EN ISO 15883 1 7 2 3 Manual disinfection external WARNING Incomplete disinfection Infection hazard Only use disinfection procedures that are verifie...

Page 20: ...als e g WL cid made by ALPRO Cover the medical device with the KaVo CLEANpac bag and place it on the corresponding care adapter Hold the can vertically Press the spray key for at least 3 seconds Remov...

Page 21: ...device in an ultrasonic device Malfunction and material damage Clean manually or in a washer disinfector only 7 3 1 Automated internal and external cleaning and internal and external disinfection KaVo...

Page 22: ...ds servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool from the medical device Cover the medical device...

Page 23: ...to coupling service point four on the far right A MULTIflex adaptor must be mounted there Press the handpiece together with the guide bush of the chuck to be ser viced against the tip of the service...

Page 24: ...ocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool from the medical device Service the product in the QUATTROcare See also 2 Instructions for us...

Page 25: ...erilisation parameters Select a suitable procedure depending on the available autoclave from the fol lowing sterilisation processes Steriliser with triple pre vacuum at least 3 minutes at 134 C 1 C 4...

Page 26: ...33 Replacement filter 1 002 0271 Wrench for water filter 1 002 0321 Instrument stand 2151 0 411 9501 Nozzle pin 0 410 0921 Insert for turbine handpieces 0 411 9902 Cleanpac 10 units 0 411 9691 MULTIfl...

Page 27: ...not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection inst...

Page 28: ......

Page 29: ......

Page 30: ...1 010 1312 bd 20171017 04 en...

Reviews: