Instructions for use MASTERtorque LUX M9000 L COLOR - 1.007.9900 MASTERtorque LUX M9000 L COLOR -
1.007.7100
7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing
20 / 30
▶
Follow the instructions for use of the disinfectant.
7.2.4 Manual disinfection - internal
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
NOTICE
Never disinfect the handpiece with chloride-containing products.
Malfunction and material damage.
▶
Only disinfect in a washer disinfector or manually.
The efficacy of manual internal disinfection must be demonstrated by the man-
ufacturer of the disinfection agent. With KaVo products, use only disinfection
agents that have been released by KaVo with respect to the compatibility of
materials (e.g. WL-cid / made by ALPRO).
▶
Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.
▶
Hold the can vertically.
▶
Press the spray key for at least 3 seconds.
▶
Remove the medical device from the spray attachment and let the disinfect-
ant act for 2 minutes.
▶
Follow the instructions for use of the disinfectant.
7.2.5 Manual drying
Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.
▶
Hold the can vertically.
▶
Press the spray key for at least 3 seconds.
See also:
2
KaVo DRYspray Instructions for Use
Note
KaVo CLEANspray and KaVo DRYspray for manual interior cleaning are only
available in the following countries:
Belgium, Denmark, Germany, Finland, France, United Kingdom, Italy, Latvia,
Lithuania, Luxembourg, Netherlands, Norway, Austria, Poland, Portugal,
Sweden, Switzerland, Spain and the Czech Republic.
In other countries interior cleaning can only be carried out with washer disin-
fectors in accordance with EN ISO 15883-1.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.