4 Start up and shut down | 4.1 Mounting the MULTIflex (LUX) / MULTIflex LED coupling
4 Start up and shut down
WARNING
Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶
Prior to initial startup and after each use, reprocess the product and
accessories.
See also:
2
7 Reprocessing steps in accordance with ISO 17664, Page 18
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
▶
Reprocess and sterilise the product and accessories before disposal.
See also:
2
7 Reprocessing steps in accordance with ISO 17664, Page 18
NOTICE
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions.
▶
Make sure that the supply of cooling air is dry, clean, and uncontaminated
according to EN ISO 7494-2.
4.1 Mounting the MULTIflex (LUX) / MULTIflex LED
coupling
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked in place can release from the MUL-
TIflex (LUX) / MULTIflex LED coupling during treatment.
▶
Before each treatment, check if the medical device is securely locked onto
the MULTIflex (LUX) / MULTIflex LED coupling by pulling on it.
▶
Screw the MULTIflex (LUX) / MULTIflex LED coupling onto the turbine hose
and tighten it with the wrench
Mat. no. 0.411.1563
.
▶
Rotate the spray ring on the MULTIflex (LUX) / MULTIflex LED coupling in
order to regulate the water supply.
4.2 Checking the amount of water
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Insufficient spray water can cause the medical device to overheat and damage
the pulp and tooth.
▶
Adjust the water amount for the spray cooling to a minimum of 50 ml/min
(3.1 inch
3
).
▶
Check spray water channels and if necessary clean spray nozzles with the
nozzle needle (
Mat. no. 0.410.0921
).
▶
Check water filter and replace, if necessary.
11 / 30