Instructions for use KaVoLUX 540 LED U / T
7 Rehabilitation methods according to DIN EN ISO 17664 | 7.3 Cleaning and disinfecting the handles
7.3 Cleaning and disinfecting the handles
▶
Remove the handles by pressing the spring-loaded catches.
▶
Clean the handles under running water.
▶
Disinfect the white handles (
Mat. no. 1.001.4472
) and the grey handles
(
Mat. no. 10024489
) with a disinfectant wipe or thermodisinfection (slight
discolouration possible).
Approved disinfectants:
▪ Incidin Liquid (Ecolab)
▪ FD 322 (Dürr)
▪ Microcide AF Liquid (Schülke & Mayr)
▶
Use disinfectants in accordance with the Instructions for Use of the manu-
facturer.
See also:
2
Disinfectant material safety data sheet
▶
Snap the handles on again.
7.4 Sterilisation
7.4.1 Sterilising the handles
Note
The white handles (
Mat. no. 10014472
) cannot be sterilised.
▶
Sterilise the grey handles (
Mat. no. 10024489
) for 4 minutes at 134
o
C ±
1
o
C packed with B cycle.