Instructions for use KaVoLUX 540 LED U / T
5 Mechanical settings | 5.3 Setting the light head brakes
76 / 110
▶
Use the adjusting screw
③
to set the smooth running of the 3D joint mo-
tion.
▶
Snap the cover
⑤
into place.
5.3.2 Adjusting the light head brake
Note
The smooth running of the operating light head can be set using two brakes.
- Brake
⑦
is for the presetting and needn´t be adjusted in normal cases. To
set the brake
⑦
, the cover
⑥
must be removed. The cover
⑥
is fastened
with a snap lock and can be removed without tools.
- Brake
①
is used for the precision adjustment and can be set without the
cover having to be removed. The adjustment can be carried out by the med-
ical staff.
Presetting
▶
Release cover
①
and remove from the side.
▶
Pivot the light head completely downwards.
▶
Preset the cross slot adjustment screw
②
with a screwdriver.