Instructions for use KaVoLUX 540 LED U / T
4 Assembly | 4.3 Assembly of the KaVoLUX 540 LED U
42 / 110
disnf. kit
ENABLED
①
Index 1 to 39
②
Bit 0 or 1
Function key
Description
S2
Reduce index
S3
Increase index
S4
Set index to 0
S5
Change note bit
ESTETICA E50/E50 Life
Note
The note bit OP light LED must be set to "0" in service mode U04 Index 9 to
activate the KaVoLUX 540 LED U operating light on the treatment centre ES-
TETICA E50/E50 Life.
▶
Press the "Remote Control" button until "U04" is shown on the display.
▶
Press button "S3" to raise the index to 9.
▶
Press the "S5" button to set the note bit to "0".
ð
"OP light LED" is shown on the display.
▶
Press and hold down button "function level" to quit the service mode.
▶
Press the "Remote Control" button.
ESTETICA E70 from SN 20100000 and E80 from SN
10100000
Note
The note bit OP light LED must be set to "1" in service mode U04 Index 28 to
activate the KaVoLUX 540 LED U operating light on the treatment units
ESTETICA E70/E80.
▶
Press the "Remote Control" button until "U04" is shown on the display.
▶
Press button "S3" to raise the index to 28.
▶
Press the "S5" button to set the note bit to "1".
ð
"OP light LED" is shown on the display.