www.kamansensors.com
18
PART NO: 860029-001
Last Revised 01/06/15
11.2.3 Now install the second sensor of the pair (S1) in the fixture and position it to
within a few mils of the required null gap. Connect the Power/Signal line to J5 and apply
power to the system. Use the output from the system as a guide in the final positioning
of this sensor (electrical nulling). Slowly move the second sensor toward or away from
the target as necessary until the system output reads 0 VDC (ideally, 0.000 V). This
output means the sensor is positioned correctly.
11.2.4 Repeat steps 11.2.1 through 11.2.3 for sensor S4 and S2.
11.2.5 The system is now ready for use.