12.3.5 Move the target through a known displacement to its maximum range and adjust
the axis 1 gain control (Figure 14) for the desired output.
12.3.6 Return the target to the null position and check the output. You most likely will
not see 0.000 V output unless you have a very good dimensional standard and very
stable fixturing located in a controlled environment. When measuring at sub micro-inch
levels, the world becomes rubber, and the fixturing may have expanded or contracted
and you won’t see 0.000 V again.
12.3.7 After returning the target to the null position, output should be 0.000 V (±10mV or
less). If the output at null exceeds this, repeat the process. Calibration should not take
more than two iterations.
12.3.8 Repeat the process for axis 2, monitoring the output at pin 6.
12.3.9
The system is ready for use.