15
6-5 Finishing Measurement
When all measurements are completed, remove the suction nozzle, and turn off the
power switch.For details on daily maintenance and storage, please refer page 23 "Daily
Maintenance and Storage".
※ When using the data collection function, please turn off the data collection
before turning off the power. If the power is turned off without turning off the
data collection, the data within the last (400) counts will be lost without stored.
※ If you had measured a strong odor like exceeding 1,000 that causes
deterioration of the sensor (refer to page 10), please execute an idling
operation under clean air environment before turning off the power. And turn
off the power after the measured value becomes stable at 300 or less.Turning
the power off while the sensor is contaminated may accelerate the deterioration
of the sensor.
※ In the case of the measured value exceeding 2,000, the measurement is not
recommended since it may lead to equipment failure or sensor sensitivity
characteristics change.
Please stop the measurement immediately in this case.
Please note that the suction nozzle is designed
to be enough hard to be pulled out.
When removing the suction nozzle,
push two points
of the black insertion port (open ring) and pull out
the nozzle.
It may not be pulled out if you push only one point.
Also, it may not be pulled out if you push it with
pulling the nozzle.
If the nozzle is difficult to be pulled out, it can be
pulled out relatively smoothly if you push two points
of insertion part of the nozzle once and then pull out.
Please be careful not to pull out by force, as it may
cause damage to the body and suction nozzle.