24
10
Error Indication
The errors displayed on the screen are as follows.
Blink:
The cumulative usage time has exceeded 3500 hours. It will need
maintenance soon.
Lights up:
The cumulative usage time has exceeded 4000 hours. Maintenance is
required.
Lights up +C001:
The cumulative usage time has exceeded 4500 hours. No more
measurement possible. An immediate maintenance is required.
Lights up +200 (Blink):
This is displayed in zero adjustment mode.(→ P.21)
Lights up + C002 / 200 (Alternate blink):
This is displayed in zero adjustment mode.The suctioned air is dirty or
the sensor is moderately degraded. Although measurement is possible,
maintenance is recommended.
Lights up +C002:
This is displayed in zero adjustment mode.The suctioned air is dirty or the
sensor is severely degraded. Please try to replace the activated carbon
cartridge to recover. If still not recover, maintenance is required.
E001
Sensor error. If the error cannot be resolved even after turning the power
off and on, maintenance is required.
E002
Pump error. If the error cannot be resolved even after turning the power
off and on, maintenance is required.
Blink:
Reading microSD. Normally, it fl ashes when writing data, but if it continues
to blink, a card read error may have occurred. Please stop measurement
and reinsert the microSD card.
Measuring
value Blink
Over range. The measured value exceeds the measurable range (more
than 2,000).
In the case of strong odor with a measured value of 2,000 or more, we do
not recommend measurement since it may lead to product malfunction.
Please stop the measurement immediately.