H2 Stationary GrainGage User’s Guide
38
H2 Stationary GrainGage Diagnostics and Alerts
The icon panel across the top of the Diagnostics dialog box allows you to access each diagnostic category.
All Diagnostics screens show the
Tare
and
Enter Glean
(or
Exit Glean
) buttons, as explained in the
following table:
Diagnostics Tare and Enter/Exit Glean
Item
Description
Tare button
Use the Tare button to zero all load cells and the moisture sensor . The tare process
is as follows:
•
Check the GrainGage and make sure it is free of grain or debris .
•
Tap
Tare
.
•
The gates open and clear the GrainGage .
•
The gates close .
•
Load cell voltages average over 4 seconds .
•
Mirus then uses that average to establish the zero weight readings for the
empty weigh bucket .
•
Mirus also resets the moisture to zero .
Enter Glean
button
This function is not used with the H2 Stationary GrainGage
.
Use the
Enter Glean
button to put the GrainGage in Glean Mode . This mode
opens all gates in the GrainGage so that grain flows straight through unimpeded .
Use Glean Mode when you need to glean off border plots . In this mode, the
button changes to
Exit Glean
.
No data will be collected or recorded in Glean
Mode.
Exit Glean
button
This function is not used with the H2 Stationary GrainGage
.
Use the
Exit Glean
button to put the GrainGage into Harvest Mode and return to
measuring the grain and recording data .
3.1.1 Tare the GrainGage
In the Diagnostics box,
1. Tap
Tare
.
2. After the tare is complete, tap the check icon to exit
Diagnostics.
Note: Taring should be the last thing you do before you
start to harvest.
Summary of Contents for HarvestMaster GRAINGAGE H2 STATIONARY
Page 1: ...USER S MANUAL...
Page 7: ...CHAPTER ONE Introduc tion...
Page 18: ...CHAPTER T WO Mirus and H2 Stationar y GrainGage Setup...
Page 36: ...CHAPTER THREE H2 Stationar y GrainGage Diagnos tic s and Aler t s...
Page 44: ...CHAPTER FOUR H2 Stationar y GrainGage Calibration...
Page 60: ...CHAPTER FIVE Field Maps...
Page 69: ...CHAPTER SIX Har ves t Mode...
Page 87: ...Standard Grain Information APPENDIX A...
Page 89: ...Troubleshooting Mirus APPENDIX B...
Page 94: ...Limited Warrant y APPENDIX C...