8 Programming
62
8.4 Configuration level 1
At the configuration level 1 it is possible to
configure channel-specific parameters.
From the parameter level, the configuration
level 1 is reached by simultaneously press-
ing keys
h
+
R
, from the configuration
level 2 by pressing the
e
key.
The parameter level 1 covers the following
parameters:
- writing status
- measurement/signal input
- scaling
- channel designation
- limit operation
- plotarea (zoom)
- offset (presentation range)
Following the input of a correct
codenumber when calling up the pa-
rameter level, the recorder interrupts
signal acquisition and recording at
the configuration level 1. The param-
eters can be viewed and altered.
Signal acquisition, processing and
recording are interrupted.
The relays of the external relay mod-
ule ER8 remain in their current sta-
tus. They are only operated again
when the recorder is at the operating
level or in the basic status.
Display
Edit
Selection/input
with keys
Enter
Continue with
C1-LEVEL
h
forwards
R
backwards