8 Programming
53
8.2.4 Level inhibit and code request
The parameter level is protected by a 4-digit
codenumber.
Two codenumbers are available:
- codenumber for a limited set of parameters
(factory-set to 9200)
- codenumber for a full set of parameters
(factory-set to 9210)
Which parameters belong to the limited set
of parameters is defined in the set-up pro-
gram under Edit
Instrument operation.
Parameters which are not required for the
applications or which are only programmed
once can be marked there accordingly. Af-
ter the new set-up has been transferred to
the pen recorder, they are not indicated any
more on the instrument, .
If one of the two codenumbers is entered,
the following are interrupted:
- signal acquisition, and
- recording
Additionally:
- events are no longer acquired,
- reports are aborted and reset,
- limit values are no longer monitored, and
- the outputs remain in their latest status.
The parameters can now be programmed.
Summary of Contents for LOGOLINE 500d
Page 106: ...9 Consumables 106 9 2 Removing and replacing the chart cassette S F ENTER E S F ENTER E...
Page 108: ...9 Consumables 108 9 2 1 Changing the roll chart 1 5 2 6 3 7 4 8...
Page 109: ...9 Consumables 109 9 2 2 Changing the fanfold chart 1 4 2 5 3...
Page 110: ...110 10 Extra Codes and accessories 10 1 Converting the chart cassette 1 3 2 4...
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......