9 Calibrating a sensor with a standard signal
64
ductivity.
✱
Start the calibration, see "Ways to start the calibration", page 54.
✱
Select REL. CELL CONST.
✱
Press the
key.
✱
When the measured value is stable, press the
key
✱
The measured conductivity value flashes on the display.
✱
Use the
or
keys to set the value to the actual conductivity.
✱
Press the
key;
the relative cell constant determined by the instrument is displayed (as a
%).
✱
Use the
key to accept the temperature coefficient or
the
key to reject it.
The current measurement value and the temperature are displayed.
PGM
PGM
PGM
PGM
EXIT