57
9 Calibrating a sensor with a standard signal
✱
Set the displayed value to the required value (usually 0) with the
and
keys; then press
to continue.
✱
Now bring the system to a second defined state (e.g. when measuring
liquid level, container full).
Wait until the display value has stabilized; then press
to
continue
✱
Set the displayed value to "Maximum" (usually 100%) with the
and
keys; then press
to continue.
The zero point and slope determined by the instrument are displayed.
✱
Use the
key to accept the calibrated values or
reject them with the
key.
✱
The instrument returns to Measuring mode.
PGM
PGM
PGM
PGM
EXIT