105
16 Appendix
16.2 Parameters of the User level
When there are numerous instrument parameters to configure, it is advisable
to make a note in the table below of all the parameters to be changed and to
work through these parameters in the given order.
The following list shows the maximum number of parameters that
can be modified.
Some of these parameters will not be visible (and therefore not
editable) for your particular instrument, depending on the
configuration.
Parameter
Selection/value range
Factory setting
New setting
Conductivity input
Cell constant
0.01/0.1/0.5/1.0/3.0/10.0
Relative cell constant
and
Relative cell constant MB 2
20.0 to 100.0 to 500.0
Offset
and
offset MB 2
-20.00 to 0.00 to 20.00 % of the display range
Temperature
compensation
and
temperature compensation
MB 2
None
Linear
Natural waters
ASTM 1125 neutral
ASTM 1125 acidic
ASTM 1125 alkaline
Temperature
compensation source
Temperature input
Option input 1
Option input 2
Option input 3
Manual temperature input
Temperature coefficient
and
temperature coefficient
MB 2
0.00 to 2.20 to 8.00 %/K
Reference temperature
15.0 to 25.0 to 35.0 °C
Pollution recognition
Off
On
Broken sensor detection
Off
On
Filter time constant
0.0 to 2.0 to 25.0 seconds
Calibration interval
0 to 99 days (0 = timer not active)
Differential measurement Off
Main input - (minus) Option input 1
Main input - (minus) Option input 2
Main input - (minus) Option input 3
Option input 1 - (minus) Main input
Option input 2 - (minus) Main input
Option input 3 - (minus) Main input