9 Calibrating a sensor with a standard signal
✱
Set the displayed value to the second buffer solution value with the
or
keys; then press
to continue.
The zero point and slope determined by the instrument are displayed.
✱
Use the
key to accept the calibrated values or
reject them with the
key.
The instrument returns to Measuring mode.
9.4
Conductivity operating mode
9.4.1
Calibration of the relative cell constant
The transmitter is in "Measuring mode".
✱
Immerse the conductivity sensor in a reference solution with a known con-
PGM
PGM
EXIT
This example is based on a conductivity sensor with a connected two-wire
transmitter.