43
7 Commissioning
7.2.2
Measurement of ultra-pure water with 2-electrode measuring
sensor
Layout
Electrical connection
See chapter 5 "Installation", page 14.
Task
Measurement range:
0 to 2.00 µS/cm
Cell constant K:
0.01 1/cm
Output signal:
4 to 20 mA
Temperature measurement Pt100
Limit monitoring:
Limit value function
Limit value 1:
USP
USP limit monitoring
Data sheet
(1)
Transmitter/controller type 202552
202552
(2)
Conductivity sensor on the main board
202924
(3)
Conductivity cable
202990
(1)
(2)
(3)
Summary of Contents for dTRANS CR 02