53
11 Instrument description
Accuracy
Relative pressure
Nominal (rated)
measuring range
0.4
1
4
10
25
60
bar
Linearity
1
0.15
0.15
0.1
0.1
0.1
0.1
% of end val. of
nom. mess. range
Accuracy at
+20°C
2
0.35
0.3
0.25
0.25
0.25
0.25
% of end val. of
nom. mess. range
Accuracy at
-20...+75°C
3
0.7
0.6
0.5
0.5
0.5
0.5
% of end val. of
nom. mess. range
Nominal (rated)
measuring range
-0.4...0.4 -1...1
-1...3
-1...9
-1...24
bar
Linearity
1
0.15
0.15
0.1
0.1
0.1
% of end val. of
nom. mess. range
Accuracy at
+20°C
2
0.35
0.3
0.25
0.25
0.25
% of end val. of
nom. mess. range
Accuracy at
-20...+75°C
3
0.7
0.6
0.5
0.5
0.5
% of end val. of
nom. mess. range
Absolute pressure
Nominal (rated)
measuring range
0.4
1
4
10
25
60
bar
Linearity
1
0.15
0.15
0.1
0.1
0.1
0.1
% of end val. of
nom. mess. range
Accuracy at
+20°C
2
0.35
0.3
0.25
0.25
0.25
0.25
% of end val. of
nom. mess. range
Accuracy at
-20...+75°C
3
0.7
0.6
0.5
0.5
0.5
0.5
% of end val. of
nom. mess. range
Long-term stability 0.2% with reference conditions to EN 61298-1
1
Linearity based on limit point setting
2
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of
measuring range.
3
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of
measuring range, thermal effect on initial value of measuring range and span.