8 Calibration

8.1

Setting the zero point (offset)

8.1.1

Automatic offset adjustment

On the instrument

See section 7.5.1 "Automatic offset adjustment", page 31.

By setup program

not possible.

8.1.2

Edited offset setting

On the instrument

See section 7.5.2 "Edited offset setting", page 31.

By setup program

Connect the instrument with the PC and start the setup program,
See section 9 "Setup program", page 45 and following.

Input / offset.

Automatic offset adjustment is only possible for instruments with a
relative pressure measuring range!

Summary of Contents for 405052

Page 1: ...JUMO DELOS SI Electronic Pressure Switch with Display B 40 5052 0 Operating Manual 2009 06 15 00526948...

Page 2: ......

Page 3: ...tions 11 4 2 Color assignment of M12 x 1 round plug 11 4 3 Terminal assignment for output 470 12 4 4 Terminal assignment for output 471 12 4 5 Terminal assignment for outputs 475 476 and 477 13 5 Moun...

Page 4: ...ting the switching point B SP 39 7 12 Setting the reset point B RSP 40 7 13 Setting the switching difference hysteresis B HYS 40 7 14 Setting the switching delay B DLY 41 7 15 Setting the display alig...

Page 5: ...cial care is required when handling components liable to damage through electrostatic discharge Note This symbol is used to draw your special attention to a remark abc1 Footnote Footnotes are remarks...

Page 6: ...g output 1 analog output 4 20 mA 1 1 PNP switching output 1 analog output 0 20 mA 1 1 PNP switching output 1 analog output 0 10 V 1 The instrument is also available in a design for use at elevated med...

Page 7: ...N 12345678 1 2 3 4 5 6 7 1 2 3 4 Type Part number Measuring range Power supply voltage 5 6 7 Output signal Manufacturing number Date of manufacture year and calendar week Type 405052 Pressure Power su...

Page 8: ...ar relative pressure 449 1 1 bar relative pressure 481 1 3 bar relative pressure 483 1 9 bar relative pressure 485 1 24 bar relative pressure Measuring range absolute pressure 486 0 400 mbar absolute...

Page 9: ...12 Clamp to DIN 32 676 DN204 613 Clamp to DIN 32 676 DN254 616 Clamp to DIN 32 676 DN504 623 Small flange connection4 652 Tank connection with groove union nut4 DN25 997 JUMO PEKA EHEDG certified 3 99...

Page 10: ...2 x 1 with 2 m PVC cable 40 00409334 4 pin angle box M 12 x1 with 2 m PVC cable 40 00522384 Setup program 40 00507861 Connecting cable only required for programming with the setup program 70 00456352...

Page 11: ...with an explosion hazard Apart from faulty installation incorrect settings on the instrument may also affect the proper functioning of the subsequent process or lead to damage You should therefore al...

Page 12: ...PNP switching output Power supply voltage 1 L 12 30 V DC 3 L GND Output 4 K1 Highside Open Collector maximum 0 25 A 2 nc 5 Interface 1 2 3 4 5 L L K1 Two PNP switching outputs Power supply voltage 1...

Page 13: ...nal assignment for outputs 475 476 and 477 One PNP switching output one analog output Power supply voltage 1 L 12 30 V DC 3 L GND Output 4 K1 Highside Open Collector maximum 0 25 A 2 Analog 0 4 20 mA...

Page 14: ...ection 11 1 Technical data page 49 Installation position The instrument can be mounted in any position The vertical installation position is recommended see illustration 5 1 1 Rotating the display The...

Page 15: ...15 5 Mounting 5 1 2 Rotating the housing The instrument housing can be rotated a maximum of 320 with the combination tool 1 320 max 1...

Page 16: ...ns of electronic pressure switches 35 42 75 M12x1 40 30 Type 405052 000 Type 405052 004 The overall height is 40 mm greater for instruments with basic type extension 004 for increased medium temperatu...

Page 17: ...connections not front flush 504 511 521 523 A Profile seal DN G3 4 B Profile seal DN G1 2 E SW 27 F SW 32 E G1 2 G1 2 1 6 7 20 31 15 19 E 11 07 26 1 4 18 NPT 1 4 18 NPT 10 4 15 A E 19 G1 4 23 5 1 5 12...

Page 18: ...flush 571 575 576 A profile seal DN G3 4 B profile seal DN G1 C O ring 26 7 x 1 78 D Drill out after tapping F SW 32 G SW 41 A F 16 16 32 G3 4 G3 4 1 6 A D F 16 24 34 20 5 G3 4 9 G3 4 24 5 1 6 1 6 20...

Page 19: ...6 5 30 RD 44 x 1 6 54 13 604 25 44 35 RD 52 x 1 6 63 15 606 40 56 48 RD 65 x 1 6 78 607 50 68 5 61 RD 78 x 1 6 92 16 Process connection DN DIN 32676 DN inches Nominal Size ISO 2852 D1 D2 612 20 15 12...

Page 20: ...ting 20 623 652 997 E SW 27 Process connection 997 is EHEDG certified For detailed information about this process connection system see data sheet 40 9711 15 3 19 26 40 RD52 x 1 6 35 45 55 63 21 31 E...

Page 21: ...e protective screw 1 Turn push the control element 2 with the enclosed combination tool or a 0 5x3 screwdriver 6 2 LC display 6 2 1 Measurement mode normal display Example The display is lit yellow 1...

Page 22: ...s No Cancel Press the combination tool more than 3 seconds 3 s Timeout No activity for more than 60 seconds 60 s To return to measuring mode No Cancel or Wait for timeout no activity performed for 60...

Page 23: ...o point offset Off P page 31 Damping filter time constant 0 00 99 99 s Parameter Display Setting range 1 Signal type for analog output 4 20 mA 0 20 mA 0 10 V Scaling start for analog output 0 00 75 00...

Page 24: ...sed only with window switching functions Switching delay for switching output only 0 00 99 99 s See section 7 10 Setting the switching function B Fct page 37 Parameter Display Setting range 1 Switchin...

Page 25: ...y for analog output only Pressure unit see parameter Uni P Percentage of the scaled range See section 7 16 Setting the display unit D Uni page 42 Version D Software version of the operating device See...

Page 26: ...the output signal See section 7 7 Setting the output signal S TyP page 32 Adjusting the scaling of the output signal restricting the measuring range See section 7 8 Setting scaling page 33 Setting the...

Page 27: ...the instrument is unprotected Unlocking Unscrew the protective screw 1 Continue briefly pressing the combination tool 2 until the third 0 from the left is flashing The color of the display also chang...

Page 28: ...for longer than 60 seconds 7 4 Selecting the unit of the measured value Uni P Unlock the instrument See section 7 2 Unlocking the instrument code entry page 27 Rotate until the bottom line shows Uni P...

Page 29: ...Rotate The measured pressure is shown in Kilopascal Rotate The measured pressure is shown in Megapascal MPa Rotate The measured pressure is shown in psi To confirm setting Press until the display is...

Page 30: ...ar 1 000 3 000 kPa 100 0 300 0 MPa 0 100 0 300 psi 14 50 43 51 0 60 bar mbar 0000 9999 bar 00 00 60 00 kPa 0000 6000 MPa 0 000 6 000 psi 000 0 870 2 1 9 bar mbar 1000 9000 bar 1 000 9 000 kPa 100 0 90...

Page 31: ...The current measured value is accepted as the zero point 7 5 2 Edited offset setting This setting is used to increase or reduce the measured pressure selectively by an adjustable value Unlock the inst...

Page 32: ...er The value is entered in seconds with two places after the decimal Unlock the instrument See section 7 2 Unlocking the instrument code entry page 27 Rotate until the bottom line shows Dam P Press 7...

Page 33: ...aling of the instrument s output signal describes how the measured pressure is converted into an output signal Simple example Actual The instrument has a nominal measuring range 1 from 0 to 4 bar and...

Page 34: ...m 4 to 20 mA 100 The scaling is 1 4 25 to 100 Inversion of the output signal The instrument provides the option of inverting the output signal 3 The output signal 0 to 20 mA becomes output signal 20 t...

Page 35: ...mbar the instrument reports an error value below lower measuring range limit and makes the corresponding error signal 0 mA available at the output At a pressure of 0 mbar the instrument makes 0 mA av...

Page 36: ...ows Sc Hi Press 7 9 Setting the error signal S Err Setting Unlock the instrument See section 7 2 Unlocking the instrument code entry page 27 Rotate until the bottom line shows S Err Press The output s...

Page 37: ...ignal 10 7 V for measuring range 0 10 V 7 10 Setting the switching function B Fct General The response of the instrument s switching output can be selected Hysteresis make contact Hysteresis break con...

Page 38: ...ct switching difference factory setting 1 Hysteresis of break contact switching difference min contact 7 10 2 Window Relay response B RSp 1 p B Sp 0 B RSp 1 p B Sp 0 1 SP Switching point 4 no SPST mak...

Page 39: ...on B Fct page 37 Setting range 0 to 100 of the nominal measuring range Factory setting 50 of the nominal measuring range Setting Unlock the instrument See section 7 2 Unlocking the instrument code ent...

Page 40: ...7 Rotate until the bottom line shows B RSP Press 7 13 Setting the switching difference hysteresis B HYS See section 7 10 Setting the switching function B Fct page 37 Setting range 0 to 100 of the nomi...

Page 41: ...try page 27 Rotate until the bottom line shows B DLY Press 7 15 Setting the display alignment D Dir Setting range std standard instrument upright turn turned instrument overhead Factory setting std Se...

Page 42: ...Unlocking the instrument code entry page 27 Rotate until the bottom line shows D Uni Press Uni P the measured value is displayed in the unit that was selected See section 7 4 Selecting the unit of the...

Page 43: ...ion 7 2 Unlocking the instrument code entry page 27 Rotate until the bottom line shows SW Di Press 7 18 Displaying the version of the signal stage software SW Si Setting range Read only Factory settin...

Page 44: ...up program not possible 8 1 2 Edited offset setting On the instrument See section 7 5 2 Edited offset setting page 31 By setup program Connect the instrument with the PC and start the setup program Se...

Page 45: ...outputs K1 and K2 Setting special functions for example tables for special linearizations Connection Data can be transferred from or to the transmitter if it is connected to the power supply See sect...

Page 46: ...o the connecting cable b and PC interface line gray c Connect the PC interface line gray c to the notebook 4 Connect the cable a to the power supply 1 and connecting cable b 9 2 Start the setup progra...

Page 47: ...47 9 Setup program HP 1...

Page 48: ...C 8 V outside of working range 6 to 8 Internal communica tion error 9 Invalid configuration 1 6 7 8 Call Customer Service see the back of the Operating Manual 2 Check the ambient temperature Check out...

Page 49: ...sic type extension 004 for elevated medium temperature Any Device standing vertically process connection on bottom Zero point correction possible locally or via setup 20 of nominal measuring range 1 m...

Page 50: ...log output Freely configurable Jump response T90 4 20 mA 1 x PNP switching output standard 0 20 mA 1 x PNP switching output 0 10 V 1 x PNP switching output 100 ms Switching output Number Switching typ...

Page 51: ...terial Surface Process seal Process connection 521 523 571 575 576 652 Process connection 997 JUMO PEKA Stainless steel 316L Ra 0 8 m All flange connections are welded and therefore have no seals FPM...

Page 52: ...85 C Permissible relative humidity in operation in storage 100 incl condensation on instrument outer sleeve 90 without condensation Permissible mechanical loading Vibration strength Shock resistance...

Page 53: ...y at 20 75 C3 0 7 0 6 0 5 0 5 0 5 of end val of nom mess range Absolute pressure Nominal rated measuring range 0 4 1 4 10 25 60 bar Linearity1 0 15 0 15 0 1 0 1 0 1 0 1 of end val of nom mess range Ac...

Page 54: ...ss 36035 Fulda Germany Phone 49 661 6003 0 Fax 49 661 6003 607 E mail mail jumo net Internet www jumo net JUMO House Temple Bank Riverway Harlow Essex CM20 2DY UK Phone 44 1279 63 55 33 Fax 44 1279 63...

Reviews: