MV / MW
23
11.
Electrical connections
/ ALARM connector
The "
ALARM
" connector may be used as output for
alarm messages.
Circuit:
Operation
= relay powered
Alarm
= relay not powered
Pin assignment:
Pin 1:
+24 V (
max. current
25 mA)
Pin 2:
0 V
Pin 3:
Alarm relay
Pin 4:
Reserved - do not use!
Pin 5:
Cooling pulse
RS232C serial interface
This port can be used to connect a computer with an
RS232C cable for remote control of the circulator.
RS232C
Pin assignments:
Pin 2
RxD
Receive Data
Pin 3
TxD
Transmit Data
Pin 5
0 VD
Signal GND
Pin 6
DTR
Data terminal ready
Pin 7
RTS
Request to send
Pin 8
CTS
Clear to send
Interface correspondence:
Circulator
Computer
Circulator
Computer
9-pole 25-pole
9-pole 9-pole
Pin 2 RxD
⇔
Pin 2 TxD
Pin 2 RxD
⇔
Pin 3 TxD
Pin 3 TxD
⇔
Pin 3 RxD
Pin 3 TxD
⇔
Pin 2 RxD
Pin 5 GND
⇔
Pin 7 GND
Pin 5 GND
⇔
Pin 5 GND
Pin 6 DTR
⇔
Pin 6 DSR
Pin 6 DTR
⇔
Pin 6 DSR
Pin 7 RTS
⇔
Pin 5 CTS
Pin 7 RTS
⇔
Pin 8 CTS
Pin 8 CTS
⇔
Pin 4 RTS
Pin 8 CTS
⇔
Pin 7 RTS
Use shielded cables only.