Operating procedures
16
6.4.
Warning functions
))
))
))
More protection for your samples in the bath!
An audible signal sounds in intervals when the actual
temperature value exceeds one of the set limits
(patented).
Setting the high temperature limit:
Press the key
.
The indicator light
blinks
and the value previously set
appears on the MULTI-DISPLAY (LED).
Use the cursor keys
to move left or right
on the MULTI-DISPLAY (LED) until the numeral
you wish to change is blinking.
Use the edit keys
to increase or decrease
the numeral value (-, 0, 1, 2, 3, ... 9).
Press enter
to store the value
(example: 39.0 °C).
Setting the low temperature limit:
Press the key
.
Follow the instructions
for
(example: 35.0 °C).
Note:
The warning functions will only be triggered when the
actual bath temperature, after start from the „OFF“ or „rOFF“
mode, lies within the set limits for 3 seconds.