Appendix
90
Auxiliary
parameters
Parameter
Setting
out_hil_00
-xxx
Setting of the actuating variable limit for cooling
capacity (0-100%)
out_hil_01
xxx
Setting of the actuating variable limit for heating
capacity (0-100%)
ATC OUT command Parameter
Setting
ATC:INT:STATUS
1 ... 10
Sets the number of correction points used for internal
adjustment. 0 = ATC_OFF
Example:
ATC:INT:STATUS˽1
A correction point is used.
ATC:INT:POINTx
˽yy.yy;zz.zz
x = 1 ... 10
yy.yy = circulator
temperature
zz.zz = correction
value
Example:
Command:
ATC:INT:POINT1˽20.00;20.10
Defines correction point 1 for internal adjustment,
with circulator temperature = 20.00 °C and
correction value = 20.01 °C.
ATC:EXT:STATUS
1 ... 10
Sets the number of correction points used for
external adjustment. 0 = ATC_OFF
Example:
ATC:EXT:STATUS˽1
A correction point is used.
ATC:EXT:POINTx
˽yy.yy;zz.zz
x = 1 ... 10
yy.yy = circulator
temperature
zz.zz = correction
value
Example:
Command:
ATC:EXT:POINT1˽20.00;20.10
Defines correction point 1, with circulator
temperature = 20.00 °C and correction value = 20.01
°C.