![Julabo Magio MX Original Operating Manual Download Page 72](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737072.webp)
Operation
72
8.10 Configure signal outputs
8.10.1 Set up Stakei outlet
A solenoid valve connected to the Stakei output can be assigned various
functions depending on the application.
The device is switched on.
A solenoid valve is connected to the Stakei output.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Connect unit>
then menu option
<Interfaces>
.
3. Use the
[Control output]
button to select the desired function for the
connected solenoid valve.
The Stakei output is set up and active.
8.10.2 Configure signal outputs of the REG/EPROG plug
The REG/EPROG socket provides three channels as signal outputs. The voltage
or current signals emitted via these channels can be externally processed.
The following values can be output:
• Setpoint: Specified setpoint
• Internal: Internal temperature sensor value
• External: External Pt100 temperature sensor value
• Performance: Value of the variable
The device is switched on.
The electronics module with analog connections is mounted.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Connect unit>
.
3. From the submenu
<Interfaces>
, select the menu option
<Analog
module>
.
4. For
[Channel 1]
select the output value setpoint, internal, external or
capacity.
5. Enter the lowest output value as the 0 V value.
6. Enter the highest output value as the 10 V value.
7. Repeat the steps 4 to 6 for
[Channel 2]
.
[Channel 3]
configures the current output. 0 mA or 4 mA can be selected
for the lowest output value.
8. Enter the lowest output value as the 0 mA/4 mA value.
9. Enter the highest output value for the 20 mA value.
The signal outputs of the REG/EPROG socket are configured.
Summary of Contents for Magio MX
Page 100: ......