![Julabo Magio MX Original Operating Manual Download Page 58](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737058.webp)
Operation
58
4. Scroll down the display to Blackbox and press the
[File]
button to change
the displayed filename if required.
5. Press the button
[Save]
.
The blackbox data is saved onto the USB stick as a .txt file.
8.5.3 Displaying alarm memory
Alarm messages are stored with date, time, alarm code and unit identifier. The
data can be read out via the menu.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Record data>
then the menu option
<Alarm
memory>
.
The device lists the stored alarm messages. Individual alarm messages can
be called up by tapping.
Alarm messages are cleared by swiping to the left and confirming the
displayed delete icon.
The alarm memory is displayed.
Summary of Contents for Magio MX