Maintenance
79
9.5 Emptying
The device must be completely drained if it is to be sent in for technical service
or is to be properly disposed of.
In general, the device should be completely emptied before longer shutdowns or
when there is a change to the external application.
CAUTION
Risk of burns due to hot bath fluid!
Bath fluid can become very hot during a temperature control process.
Contact with hot bath fluid can cause burns.
• Before draining the device, temper it to room temperature
• Avoid direct contact with hot bath fluid
• Wear protective gloves
The device is tempered to room temperature and switched off.
1. Place an adequately large collection vessel under the drainage valve.
2. Take off the bath lid.
3. Open the drain valve.
The bath fluid drains out of the bath tank into the collection vessel
provided.
4. Once the bath tank is completely drained, close the bath opening.
5. Close the drain valve.
The device is emptied. If an external system is connected, it can now be
disconnected from the device and also drained.
9.6 Clean device
The circulator and bath tank, and also a cooling machine if connected, should
be cleaned from time to time.
In addition to this, the device must be appropriately decontaminated if
hazardous substances have been spilled on or into the device.
The device is switched off and disconnected from the mains voltage.
1. Allow the device to cool down to room temperature.
2. Completely drain the bath fluid.
3. Clean the surface of the circulator and the bath tank with a damp cloth.
Some dish detergent may also be used for cleaning. If in doubt, ask
technical service for alternative cleaning mediums.
The device has now been cleaned.