Temperature System Magnum 91
33
7.3.2. Warning
functions
))
))
))
The high and low temperature warning functions accompany the working
temperature value. An audible signal sounds in intervals when the actual
temperature exceeds one of the set limits (patented).
The corresponding message appears in line 4 on the DIALOG-DISPLAY
(LCD).
Setp.: 120.00°C
>Overt.:130.00°C
Subtmp:-55.00°C
St.Pump: 2
>Overt: Setting the high temperature limit
Setting in line 2.
The value previously set appears on the DIALOG-DISPLAY (LCD).
A flashing segment indicates that a value needs to be entered.
Use the keypad to enter the new value
(examples: 130.00 °C; -13.00 °C).
Press enter
to store the selected value.
Setp.: 120.00°C
Overt.:130.00°C
>Subtmp:110.00°C
St.Pump: 2
>Subtmp:
Setting the low temperature limit
Setting in line 3.
The value previously set appears on the DIALOG-DISPLAY (LCD).
A flashing segment indicates that a value needs to be entered.
Use the keypad to enter the new value
(examples: 110.00 °C; -17.00 °C).
Press enter
to store the selected value.
Note:
The warning functions will only be triggered when the actual bath temperature, after start from
the "OFF" or "rOFF" mode, lies within the set limits for 3 seconds.
Example:
Setp: 120.00°C
IntAct 130.50°C
ExtAct: --.--°C
Control: Intern
Excess temp.
t
150
130
120
110
20
°C
The corresponding message appears in line 4 on the DIALOG-DISPLAY
(LCD).