Temperature System Magnum 91
11
2.3. Cooling
water
connection
Cooling water pressure (IN / OUT )
max.
6 bar
Difference pressure (IN - OUT )
3.5 to 6 bar
Cooling water temperature
<20 °C
Recommended quality of cooling water:
pH – value
7.5 to 9.0
Sulfate [SO4 2- ]
< 100 ppm
Hydrocarbonate [HCO3- ] / Sulphate [SO4 2-]
> 1 ppm
Hardness [Ca2+, Mg2+] / [HCO3-]
> 0,5 dH
Alkalinity
60 ppm < [HCO3-] < 300 ppm
Conductivity
< 500
μ
s / cm
Chloride (CL-)
< 50 ppm
Phosphate (PO43-)
< 2 ppm
Ammonia (NH3)
< 0.5 ppm
Free Chlorine
< 0.5 ppm
Ferri Ions (Fe3+ )
< 0.5 ppm
Mangano Ions (Mn2+)
< 0.05 ppm
Carbon dioxide (CO2)
< 10 ppm
Hydrosulfide (H2S)
< 50 ppm
Content of oxygen
< 0.1 ppm
Algae growth
impermissible
Suspended solids
impermissible
Notice:
Danger of corrosion of heat exchanger due to unsuitable quality of cooling water.
•
Due to its high content of lime hart water is not suitable for cooling and causes calcination
of the heat exchanger.
•
Ferrous water or water containing ferrous particles will cause formation of rust even in
heat exchangers made of stainless steel.
•
Chlorous water will cause pitting corrosion in heat exchangers made of stainless steel.
•
Due to its corrosive characteristics distilled and deionized water is unsuitable and will
cause corrosion of the bath. .
•
Due to its corrosive characteristics sea water is not suitable.
•
Due to its microbiological (bacteria) components which settle in the heat exchanger
untreated and unpurified river water and water from cooling towers is unsuitable.
•
Avoid particulate matter in cooling water.
•
Avoid
putrid
water.
Notice:
Cooling
water
circuit
Risk of oil leaking from the cooling circuit (compressor) of the recirculating cooler into the cooling
water in case of a fault in the circuit!
Observe the laws and regulations of the water distribution company valid in the
location where the unit is operated.