Operating procedures
30
6. Operating
procedures
6.1. Switching on / Selecting the language
JULABO
Magnum
91
V1.91-AE
Switching on:
IN
The unit is operated by turning the mains power switch (1).
The integrated pilot lamp illuminates.
L
During the self-test all segments of the VFD-Info-Display, all control
indicators and the DIALOG-DISPLAY light up.
Then the software version number (example: V1.91-AE) appears for
a short moment and the message "
OFF
" indicates the unit is ready to
operate.
L
Message >WAIT< when adjusting >Mode: sys close<:.
During the message >WAIT< the ventilation valve is opened for 10
seconds. So, the air which has eventually been collected during the
standstill, can escape.
L
The unit returns to the previous operating mode that was active
before it was turned off:
keypad control mode
(manual operation via the unit itself/RD)
or
remote control mode
(operation via PC).
Selecting the language:
There are two options for the language of the DIALOG-DISPLAY (LCD):
German or English. Select the desired language in the MENU level under
the submenu >Configuration.
Soll: 20.00°C
IntIst xxx.00°C
ExtIst:
Regelung:
>Konfiguration
Regelparameter
Konfiguration
Identif. xxxxx
Sollwert xxxxx
Autostart xxx
Standby xxxx
>Sprache xxxxxxx
Setp.: 20.00°C
IntAct xxx.xx°C
ExtAct: xxx.xx°C
Control: intern
Press the respective keys in the following order:
1. MENU key
1x
2. Enter key
1x
3. Cursor key
up to Submenu "Sprache/Language"
4. P-key
1x
5. Enter key
1x
6. Escape key
ESC
2x
The DIALOG-DISPLAY (LCD) helps to follow up the individual settings
(example: swap the language from German to English).