Remote control
62
11. Remote control
11.1. Setup for remote control
9
6
5
1
SERIAL
S xx.xx
P x.xx
R
•
Check the interface parameters for both interfaces (on circulator and PC) and
make sure they match. (Serial interface see page 52)
•
Select the "Configuration" submenu and select the option "Setpoint" to define the
interface (see page 39).
If the temperature system is put into remote control mode via the configuration
level, the display will read
"R -OFF-" = REMOTE STOP.
•
Connect both units with an interface cable.
Like all parameters which can be entered through the keypad, interface parameters
are stored in memory even after the circulator is turned off.
Factory settings:
RS232
BAUDRATE
4800 bauds
PARITY even
parity
HANDSHAKE
Protocol
RTS/CTS
(hardware
handshake)
Data
bits
7
Stop
bit
1