Temperature System Magnum 91
47
Int.Programmer
>Edit
Delete
Set Clock
>Edit Profile: 1
Step: 6
Setp: 100.00°C
Time[h.m] 01:05
Examples:
Profile No 1:
Step
Setpoint
Time
Edit
Compile profiles:
•
A flashing segment indicates that a number needs to be entered.
Under submenu "Edit Profile" enter a profile number.
Six profiles may be stored (nos. 0 to 5).
•
Then programme the desired values for each section.
Use the keypad to set section number, target temperature and time
period. Set each entry with enter
.
When the program is running, only sections having complete
information for target temperature and time period are considered.
It makes sense, to leave out section numbers in the profile, in order to
use them later for corrections in the profile.
Important:
If a time of 00:00 is set for a profile, the profile is continued with the
next section only after the setpoint temperature (±0.2 °C) is reached.
IN
Press
escape
ESC
to return to the previous menu level.
Example:
Step
(No.)
1 3 5 6 9 11 14
Setpoint (°C)
50 50 100
100 80 80 20
Time
(h.m) 00:08 00.02 00:10 01:05 00:10 00:05 00:30
20
40
60
80
100
120
140
160
180
Steps
1
3
5
6
9 11
14
10
20
30
40
50
60
120 min
Step
Edit Profile: 1
>Step: 3
Setp: 50:00°C
Time[h.m] 00:02
Display sections:
1.
Use the cursor keys
to select the submenu "Step", enter the
desired number and press enter
.
or
2.
If the submenu „Step“ is selected, section by section can be shown in
ascending order by the P-key
.
Each previously set value is displayed.
Die eingestellten Werte werden jeweils angezeigt.