Menu functions
54
8.9. Pump
Pump
>Pump Off
Mode: drain
Limits
Interface
Temp.Sensor
>Pump
In this submenu the circulation pump can be turned on separately, without
heating element or cooling unit.
The menu option „Modus“ refers to the filling valve and the ventilation
valve. Depending on the operating status (drain / fill / sys open / sys close)
the valves are switched off automatically.
•
Press
enter
to select the submenu "Pump".
•
Select the desired option with the up/down cursor keys
.
Enter the desired value for the flashing digit.
Pump
OFF/
ON
Mode:
drain / fill / sys open / sys close
•
Press the P-key
to select the desired parameter and set with
enter
.
•
Return to the previous menu level with escape
ESC
.
Explication of terms:
>Mode: System closed<
stands for closed, external system.
(
Magnum
principle page 16)
In this operating status the filling valve is opened and the ventilation valve is closed.
>Mode: Sys open<
stands for open external system
(not recommended by JULABO)
>Mode: drain<
. (Chapter „Draining“ see page 29)
In this operating status the filling valve and the ventilation valve are opened.
In the configuration of this unit the >Mode: fill< means the same as >Mode: drain<.