38 DePuy Synthes User Manual Small Battery Drive II
9.
Automated cleaning cycle parameters
Note:
Automated washing shall be conducted in a
validated washer-disinfector in compliance to ISO
15883-1 and-2, or to an equivalent standard. Load
the device components in the washer-disinfector in
accordance with manufacturer's instructions,
ensuring that the devices and lumens can drain
freely. Automated washing can be included as part
of a validated washing, disinfection, and/or drying
cycle in accordance to manufacturer's instructions.
Cycle
Minimum
Time (minutes)
Minimum
Temperature / Water
Pre-wash
2
Cold tap water
Wash I
2
Cold tap water (< 40 °C)
Wash II
5
Warm tap water (> 40 °C)
Rinse
2
Warm DI or PURW (> 40 °C)
Thermal disinfection
5
>93 °C
Dry
40
>90 °C
10.
Inspect the device.
Remove all the devices from
the washing basket. Inspect the cannulations,
coupling sleeves, etc. for visible soil. If necessary,
repeat the manual pre-clean / automated cleaning
cycle. Confirm that all parts are having a lack of
moisture both inside and outside. A reduction of the
dry time may lead to damage to the electrical
components in the power tool due to the presence
of moisture. Such damage will not be covered by
warranty.
Warning:
Mechanical cleaning is an additional
stress for power equipment, especially for seals and
bearings. Therefore, devices must be properly
lubricated after automated cleaning. Furthermore,
the device must be serviced at least once per year as
specified under the section “Repair and Technical
Services”.
Care and Maintenance
Cleaning and Disinfection
Automated Cleaning Instructions with Manual Pre-Cleaning