JK2830 Operation Manual Chapter 5 Execute LCR operation and some examples
6.6.2
The used instruments and devices
No.
Instrument and Device
Specification
1
Standard
capacitor
100pF
0.02
%
D is known
1000pF
10000pF
10nF
0.1uF
1uF
2
AC
standard resistor
10Ω
0.02
%
100Ω
1kΩ
10kΩ
100kΩ
3
DC
standard resistor
0.1Ω
0.02
%
1Ω
10Ω
100Ω
1kΩ
10kΩ
100kΩ
4
Standard inductor
100μH
0.02%
1mH
10mH
100mH
5
Frequency counter
(
0
~
1000
)
MHz
6
Digital Multimeter
0.5
%
7
Insulation resistance meter
500V 10 levels
8
Hipot Tester
0.25kW
(
0
~
500
)
V
6.6.3
Function check
Ensure function keys, displayer and terminal etc. can work normally.
6.6.4
Test signal level
Adjust multimeter in AC voltage range, where one test cable is connected to H
CUR
and the other is connected to ground terminal. Change level as: 10mV, 20mV,
100mV, 200mV, 1V, 2V, the reading should meet the demand of test signal level in
this chapter.