JK2830 Series Operation Manual Chapter 3 Introduction to [DISP]
19
3.3
<LIST SWEEP DISP>
Up to 100 test frequencies, test levels or DC bias can be set on this page. Users
can set the high and the low limits for each list-sweep test point. Auto sweep test
will be made on these test points. And their test results will be compared with
their limits.
Press down the menu key
[MEAS]
and then the soft key
LIST SWEEP
to enter
into the
<LIST SWEEP>
page, shown as below.
Test points will be automatically tested in a scanning mode. Meanwhile,
comparison will be made between test results and limit values. In the process of
list sweep test
, “
*
” denotes the current sweep test point. The following control
parameters can be set on <
LIST SWEEP
>.
◼
Sweep mode (
SWEEP MODE
)
◼
Fail mode(
FAIL MODE
)
There are 2 zones on this page:
LIST SWEEP DISP
and
mode
. List sweep points
cannot be set on this page but can be set on <
LIST SWEEP SETUP
>.
3.3.1
Sweep mode
The list sweep function of JK2830 can make automatic sweep test for up to 100
points’ test frequencies, tes
t levels or DC bias. Two sweep modes are available on
JK2830: SEQ and STEP. In SEQ mode, each press of [
TRIG
] will direct JK2830 to
automatically test all list sweep test points. In STEP mode, each press of [
TRIG
]
will direct JK2830 to test one list sweep point.