JK2830 Operation Manual Chapter 5 Execute LCR operation and some examples
6.2.3
Test signal level
Mode
Range
Accuracy
Resolution
Voltage Normal 10mV
RMS
—
2V
RMS
±
(
10%×preset value+2mV
)
10mV
6.2.4
Output impedance
30Ω, 50Ω and 100Ω are selectable.
6.2.5
Monitor for test signal level
Mode
Range
Accuracy
Voltage
10mV
RMS
—
2V
RMS
±
(
3%×0.5mV
)
Current
50μA
RMS
—
20mA
RMS
±
(
3%×r5μA
)
6.2.6
Maximum measurement display range
Parameter
Measurement display range
L
0.00001μH
~
99.9999kH
C
0.00001pF
~
9.99999F
Z
、
R
、
X
0.00001Ω
~
99.9999MΩ
Y
、
B
、
G
0.00001μs
~
99.9999S
D
0.00001
—
9.99999
Q
0.00001
—
99999.9
θ
Deg -179.999°
~
179.999°
Rad -3.14159
~
3.14159
DCR
0.1m
Ω
~
999.
999MΩ
6.3
Measurement accuracy
Test accuracy includes stability
、
temperature coefficient
、
linear degree
、
test
repeatability and calibration inter-error.
Check the accuracy of instrument should be under the following circumstances
:
a
.
warm-up time
:
≥ 30 min
b
.
cable
:
0m
,
1m
c
.
correct open and short correction after warming up
d
.
DC bias is in the position of “OFF”
e
.
The range works in “AUTO” to select correct test range