REL
The REL command will switch the meter to relative
power mode.
Interface:
GPIB and RS-‐232/USB
Syntax:
“REL”<CR>
Returned Format:
None
BRM
The BRM command will switch the meter to BR
mode.
Interface:
GPIB and RS-‐232/USB
Syntax:
“BRM”<CR>
Returned Format:
None
DUL
The DUL command will switch the meter to dual
display mode.
Interface:
GPIB and RS-‐232/USB
Syntax:
“DUL”<CR>
Returned Format:
None
MODE?
The MODE? command queries the current operation
mode, query return “ABS”, “REL”, “BRM” or “DUL”
for the four different operating modes.
Interface:
GPIB and RS-‐232/USB
Syntax:
“MODE?”<CR>
Returned Format:
<Mode><NL> where <Mode> = “ABS”, “REL”, “BRM”,
or “DUL”
SWL
The SWL command switches the meter to the next
wavelength or to a specific wavelength using the
wavelength (um) as reference.
Interface:
GPIB and RS-‐232/USB
Syntax:
“SWL”<CR> (switches to next wavelength)
“SWL x.x”<CR> (where x.x = 0.8, 1.3, 1.4, 1.5 or 1.6).
Example, “SWL 1.3” switches to 1310nm
Returned Format: